| Model Number |
30-401LITE |
| Medical Device Problem Code |
Material Disintegration (1177)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
02/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.The device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
|
| |
|
Event or Problem Description
|
|
It was reported that during a myosure lite procedure on (b)(6) 2025, small metal fragments/pieces were noted floating in the uterine cavity.The metal pieces were removed and no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device was returned for investigation: a visual inspection was conducted, and there were no signs of physical damage.Functional testing was conducted, and the device was opening and closing the window as intended.A dissection test was conducted, and the blade had scratches on it and the outer tube was bent at the tip.Hence, the complaint can be confirmed.This observation will be monitored and trended.A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.
|
| |
|
Search Alerts/Recalls
|