• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOLOGIC, INC MYOSURE; HYSTEROSCOPE (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC, INC MYOSURE; HYSTEROSCOPE (AND ACCESSORIES) Back to Search Results
Model Number 30-401LITE
Medical Device Problem Code Material Disintegration (1177)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 02/18/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.The device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
 
Event or Problem Description
It was reported that during a myosure lite procedure on (b)(6) 2025, small metal fragments/pieces were noted floating in the uterine cavity.The metal pieces were removed and no patient injury reported.
 
Additional Manufacturer Narrative
Device was returned for investigation: a visual inspection was conducted, and there were no signs of physical damage.Functional testing was conducted, and the device was opening and closing the window as intended.A dissection test was conducted, and the blade had scratches on it and the outer tube was bent at the tip.Hence, the complaint can be confirmed.This observation will be monitored and trended.A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MYOSURE
Common Device Name
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
HOLOGIC, INC
250 campus drive
marlborough MA 01752
Manufacturer (Section G)
HOLOGIC, INC.
250 campus drive
marlborough MA 01752
Manufacturer Contact
daniel guevara
562 parkway
coyol free park zone b24
alajuela 20102
MDR Report Key21635444
Report Number1222780-2025-00124
Device Sequence Number20703155
Product Code HIH
UDI-Device Identifier15420045505070
UDI-Public(01)15420045505070(17)270825(10)24J09RC
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K142029
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number30-401LITE
Device Catalogue Number30-401LITE
Device Lot Number24J09RC
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/18/2025
Supplement Report FDA Received Date04/23/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/09/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-