EXACTECH, INC. EQUINOXE GLENOID, PEGGED BETA, MEDIUM; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
|
Back to Search Results |
|
| Catalog Number |
314-02-13 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Inadequate Osseointegration (2646); Aseptic loosening (4883)
|
| Date of Event |
11/08/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(d10) concomitant devices: 300-01-13 - eq humeral stem primary, press fit 13mm: (b)(6), 300-10-15 - equinoxe replicator plate 1.5mm o/s: (b)(6), 310-02-47 - eq humeral head tall, 47mm: (beta): (b)(6).The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent an initial total shoulder replacement on the right side.Subsequently, the patient experienced aseptic glenoid loosening, failed atsa.As a result, approximately 13 years after initial replacement, the patient underwent a right shoulder revision.No further impact to the patient was reported.No other information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: g3, h6, h11.Mdr section codes updated/corrected: b, c, d, e, g.The revision reported may be the result of the glenoid loosening as reported.However, the loosening cannot be confirmed and any potential contributions from user or patient-related contributing factors to the event cannot be evaluated as no radiograph, images, or clinical information was provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|