| Model Number |
SK38 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Pain (1994); Swelling/ Edema (4577)
|
| Date of Event |
02/17/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
It was reported that after an initial bunion surgery on (b)(6)204, an sk38 screw was removed in office on (b)(6)2025 due to patient complaints of pain, swelling and a bump on the side of their foot.Upon x-ray review, it was discovered a screw was backing out.No other patient outcomes were reported as a result of this event.No devices were returned for evaluation.The device history records were reviewed and no issues were identified during the manufacture and release of the devices that could have contributed to what was reported.Although several factors can contribute to the reported event, the most likely cause cannot be determined due to the limited information provided, and no device being returned.No deficiencies or malfunctions have been alleged/found with any tmc device.The company will supplement this mdr as necessary and appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that after an initial bunion surgery on (b)(6)2024, an sk38 screw was removed in office on (b)(6)2025 due to patient complaints of pain, swelling and a bump on the side of their foot.Upon x-ray review, it was discovered a screw was backing out.No other patient outcomes were reported as a result of this event.
|
| |
|