A healthcare facility in france reported via a fisher & paykel healthcare (f&p) field representative that an opt964 optiflow + duet asymmetrical nasal cannula was found deformed and leaking during patient use.The healthcare facility stated that the subject device had been in use for a few minutes.The healthcare facility reported that the patient started to desaturate.The subject device was then replaced, and the patient recovered.There was no further patient consequence reported.Fisher & paykel healthcare (f&p) have requested further information about the reported event.However, no further information has been provided to date.F&p will provide a follow-up report upon completion of our investigation.
|
|
(b)(4).Corrected data: b5, d9, g3, g6, h2, h3, h6.Product background: the opt964 optiflow + duet asymmetrical nasal cannula is a nasal cannula patient interface for delivery of humidified respiratory gases, including those who are receiving nasal high flow therapy (nhf).The cannula consists of a lightweight delivery tube connected to a rigid plastic base and soft nasal prongs (nasal interface).This allows the device to be light weight and durable.The interface is held in place by a head strap and features a head strap clip which works in tandem with the tubing clip (attaches to the patient's clothing/bedding) to support the weight of the circuit and prevent the cannula being dislodged.Optiflow + interfaces are designed for use with the airvo series humidifiers and are also compatible with the fisher & paykel healthcare (f&p) mr850 and 950 humidification system devices.Optiflow + interfaces are intended for use with spontaneously breathing patients who would benefit from receiving high flow warmed and humidified respiratory gases.Optiflow + interfaces are not intended for life support and appropriate patient monitoring must be used at all times.Method: the complaint opt964 optiflow + duet asymmetrical nasal cannula was received at f&p in new zealand for investigation, where it was visually inspected.F&p's investigation is based on f&p's evaluation of the subject device, information provided by the healthcare facility, previous investigations of similar complaints, and f&p's knowledge of the product.Results: visual inspection of the opt964 optiflow + duet asymmetrical nasal cannula revealed that the tubing was torn and stretched in two separate places near the manifold.The film of the tubing was wrinkled which indicates the damage was consistent with an external force being applied to the tubing.The healthcare facility reported that the patient started to desaturate.The subject device was then replaced, and the patient recovered.F&p requested for further information about the reported event, such as the level and duration of the patient desaturation and the final patient outcome of the event.However, the healthcare facility stated that no further information regarding the patient's condition is available.There were no further patient consequences reported by the healthcare facility.Conclusion: f&p's investigation was unable to determine the exact cause of the observed damage to the opt964 optiflow + duet asymmetrical nasal cannula.Based on f&p's knowledge of the product the tubing was likely subjected to excessive force.The healthcare facility stated that the subject device had been in use for a few minutes.Visual inspection of the subject device found the tubing had been torn and stretched in two separate places and the film of the tubing was wrinkled.F&p's manufacturing controls for the optiflow + tubing include inspections during production for visual defects including cracks, tears, moulding defects, contamination, inclusions, discoloration and stretching or deformation.The subject opt964 optiflow + duet asymmetrical nasal cannula would have met the required specifications.The user instructions which accompany the opt964 optiflow + duet asymmetrical nasal cannula show in pictorial format the correct placement and fitting of the cannula, including ensuring the head strap clip and the tubing clip are appropriately attached.The user instructions also state: - "do not crush or stretch tube, to prevent loss of therapy." - "ensure head strap clip is attached, to prevent cannula from being pulled out of the nares." - "cannula can become unattached if not used with the head strap clip." - "attach tubing clip to clothing/bedding to prevent cannula from pulling off face." - "appropriate patient monitoring must be used at all times.Failure to monitor the patient may result in loss of therapy, serious injury or death." - "failure to use the set-up described above can compromise performance and affect patient safety.".
|
|
A healthcare facility in france reported via a fisher & paykel healthcare (f&p) field representative that an opt964 optiflow + duet asymmetrical nasal cannula was found deformed and leaking during patient use.The healthcare facility stated that the subject device had been in use for a few minutes.The healthcare facility reported that the patient started to desaturate.The subject device was then replaced, and the patient recovered.There was no further reported patient consequence.F&p requested for further information about the reported event, such as the level and duration of the patient desaturation and the final patient outcome of the event.However, the healthcare facility stated that no further information regarding the patient's condition is available.
|