| Model Number |
CYF-VHA |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/19/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during maintenance/service, the flex video scope had a disconnected distal end.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d8, h3, h6, h11.The device was returned to olympus for inspection and the reported failure was not confirmed, the distal end was loose but not disconnected.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: stress problem.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was reported that during maintenance/service, the flex video scope had a disconnected distal end.There was no patient involvement.
|
| |
|