• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRUM AB OPRA IMPLANT SYSTEM; AXOR II

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRUM AB OPRA IMPLANT SYSTEM; AXOR II Back to Search Results
Model Number 1288
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/04/2025
Type of Reportable Event Malfunction
Event or Problem Description
(b)(6) 2025: integrum has been informed that axor ii unit i9106 has fallen off from the abutment (b)(6) 2024 - (b)(6) 2025.No health consequences or patient injury reported.Manufacturing investigation performed; batch documentation reviewed.No deviations found; the device has been manufactured according to specification.(b)(6) 2025: technical investigation of unit i9106 performed.During the investigation the device functions were tested and components inspected.The reported issue could not be replicated.The device showed signs of poor maintenance.The unit has not been sent to integrum for annual service according to instructions in the ifu.The clamps were damaged and showed signs of improper donning.During the service the device was thoroughly cleaned.Damaged and worn components were replaced.Finally, the device was functionally tested with approved results.Integrum has informed the clinic on the importance of donning the axor ii according to the ifu to ensure a stable connection and prevent damage or wear to the clamps.As stated in the ifu, the device shall not be used if a stable connection cannot be achieved.Integrum has also contacted the clinic to highlight the importance of cleaning the axor ii in accordance with the ifu.Failure to adhere to the instructions may affect the axor ii's function.And remember to send the device to integrum for annual service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPRA IMPLANT SYSTEM
Common Device Name
AXOR II
Manufacturer (Section D)
INTEGRUM AB
gemenskapens gata 9
molndal, 43153
SW  43153
Manufacturer (Section G)
INTEGRUM
gemenskapens gata 9
molndal, 43153
SW   43153
Manufacturer Contact
susanne börjesson
gemenskapens gata 9
molndal, 43153
SW   43153
MDR Report Key21641543
Report Number3011386779-2025-00225
Device Sequence Number7756518
Product Code PJY
UDI-Device Identifier07340152100634
UDI-Public07340152100634
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P190009
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Remedial Action Inspection,Repair
Type of Report Initial
Report Date (Section B) 03/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number1288
Device Lot Number58766
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/03/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/03/2025
Initial Report FDA Received Date03/19/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-