• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RETRACTABLE TECHNOLOGIES, INC. VANISHPOINT®; SYRINGE, ANTISTICK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RETRACTABLE TECHNOLOGIES, INC. VANISHPOINT®; SYRINGE, ANTISTICK Back to Search Results
Model Number 10311
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/11/2025
Type of Reportable Event Malfunction
Event or Problem Description
The plunger is very difficult to push in, almost requiring 2 thumbs to push plunger to allow for the needle to retract.Coincidently, we also realized that the blue circle on these devices is also flat or somewhat concave in appearance whereas previous syringes the blue circle was convex.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VANISHPOINT®
Common Device Name
SYRINGE, ANTISTICK
Manufacturer (Section D)
RETRACTABLE TECHNOLOGIES, INC.
511 lobo ln
little elm TX 75068
MDR Report Key21641835
Report Number21641835
Device Sequence Number12678119
Product Code MEG
UDI-Device Identifier00613703103109
UDI-Public(01)00613703103109
Combination Product (Y/N)N
Initial Reporter StateVA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type User Facility
Initial Reporter Occupation Risk Manager
Type of Report Initial
Report Date (Section B) 03/11/2025
Report Date (Section F) 03/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number10311
Device Catalogue Number10311
Device Lot NumberA058B
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Event Location Hospital
Date Report to Manufacturer03/19/2025
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date03/19/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
Date Report Sent to FDA03/11/2025
-
-