H3: product analysis: evaluation could not be performed because as tool not rotate.It was noted that collet shaft assembly found corroded.This finding may have contributed to the user's experience.It was also noted depth test failed, motor shaft and winding assembly found dirty.This regulatory report is being submitted due to retrospective review through capa: 624392.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|