• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH RESTORATION ADM X3 INS 28/48; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH RESTORATION ADM X3 INS 28/48; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 7236-2-848
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Wound Dehiscence (1154); Unspecified Infection (1930)
Date of Event 02/24/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Reported event: an event regarding infection involving a adm liner was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional, and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot or sterile lot referenced.Conclusions: it was reported that the patient was revised due to infection.All stryker products sold as sterile are validated to a minimum sterility assurance level sal of 10^-6 in accordance with applicable iso standards.The exact cause of the infection event could not be determined because insufficient information was provided.Further information such as pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.The following devices were also listed in this report: delta v-40 ceramic head 28/0; cat# 6570-0-128 ; lot# 10711252, modular dual mobility insert; cat# 626-00-42e; lot# 23837253, tridentii tritanium cluster54e; cat# 702-04-54e; lot# 25025451a, exeter v40 stem 44mm no 1; cat# 0580-1-441; lot# g8757193.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.
 
Event or Problem Description
The following was reported: "bottom edge of wound started oozing.This was revision surgery.The pt presented to the emergency department with an oozing wound.She was taken to theatre that afternoon for a washout & the surgeon decided to change the head & mdm insert as well.Surgery was completed successfully.There was no delay in surgery.This was a revision procedure due to evidence of a possible infection.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESTORATION ADM X3 INS 28/48
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
anna ryan
raheen business park
limerick NA
EI   NA
61498200
MDR Report Key21642343
Report Number0002249697-2025-00311
Device Sequence Number7661795
Product Code LPH
UDI-Device Identifier07613327315974
UDI-Public07613327315974
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K182468
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number7236-2-848
Device Lot Number19895151
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 02/24/2025
Initial Report FDA Received Date03/19/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/14/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age51 YR
Patient SexFemale
Patient Weight95 KG
-
-