• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EMBECTA PENEL LIMITED - DUN LAOGHAIRE, IRELAND PEN NDL 32G 4MM PRO 90CT US; Needle, hypodermic, single lumen

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EMBECTA PENEL LIMITED - DUN LAOGHAIRE, IRELAND PEN NDL 32G 4MM PRO 90CT US; Needle, hypodermic, single lumen Back to Search Results
Model Number 320574
Medical Device Problem Code Failure to Deliver (2338)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This medwatch submission is both an initial and supplemental filing.Investigation of the results can be seen below: no samples (including photos) were returned therefore the complaint could not be confirmed and the root cause is undetermined.This is the 3rd complaint for the reported lot number.A review of the manufacturing records was performed, and no non-conformances were raised in association with this type of event for this lot.Complaints received for this device and reported condition will continue to be tracked and trended.If samples are received in the future the complaint will be reopened for further investigation.Based on the above, no additional investigation and no corrective/preventative action (capa) or situational analysis (sa) is required at this time.
 
Event or Problem Description
Consumer reported needle clog during priming.He stated that there is no insulin flow.Consumer does not re-use.Lot #: 3179396 catalog #: 320574 date of event: unknown samples: discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PEN NDL 32G 4MM PRO 90CT US
Common Device Name
Needle, hypodermic, single lumen
Manufacturer (Section D)
EMBECTA PENEL LIMITED - DUN LAOGHAIRE, IRELAND
EI 
Manufacturer Contact
avital merl
300 kimball dr.
parsippany, NJ 07054
MDR Report Key21643870
Report Number3023359743-2025-00210
Device Sequence Number7661841
Product Code FMI
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K212015
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Type of Report Initial,Followup
Report Date (Section B) 09/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number320574
Device Lot Number3179396
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/24/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/19/2025
Supplement Report FDA Received Date09/07/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
Patient RaceWhite
-
-