• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. VCL+ VIO 27IN 0 S/A CT-2; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. VCL+ VIO 27IN 0 S/A CT-2; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number VCP334H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/20/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint (b)(4).Date sent to the fda: 3/19/2025.H6 component code: g07002 - device not returned.H6 component code: c22 - photo analysis.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested, the following was obtained: - was the needle removed from the patient during the same procedure? - was there any additional tissue damage as a result of searching for the needle? - provide the source or name and title of external person providing answers to follow-up (external person submitting answers to sales rep).I was in the room during the procedure and witnessed the process to identify the location of the needle.An x-ray tech was brought into the room to x-ray the patient by placing a plate underneath the patient for the purpose of finding the needle.The needle was located and extracted from the patient through a trocar port.I did observe the needle exiting the port.There was no additional tissue damage noted or discussed after the needle was extracted - dr (b)(6) was in the room and observed the same.H3 photo evaluation: this is an analysis for a photo submitted to ethicon for evaluation.During the visual analysis, the following was observed: the photo shows a winding former with a suture that has not been dispensed, the needle is not visible in the image.Based on the photo review, no conclusion or root cause could be determined.Hands on device analysis may provide the additional evidence necessary to confirm the root cause of the reported event.Because the device was not returned our evaluation is limited.As part of ethicon quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post market surveillance.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported that a patient underwent a hysterectomy procedure on (b)(6) 2025 and suture was used.During robotic hysterectomy, the surgical tech was putting a suture down a trocar and when the suture went down the needle went missing.It was hard to find the needle.Intervention was used with x-ray, outside the patient.Near the hip.Case was completed with a different suture still an ethicon product.No patient harm was reported.Device was discarded.Additional information was requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL+ VIO 27IN 0 S/A CT-2
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key21644498
Report Number2210968-2025-02963
Device Sequence Number15357240
Product Code GAM
UDI-Device Identifier10705031052413
UDI-Public(01)10705031052413(17)280831(10)TKMKJH
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K132580
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberVCP334H
Device Lot NumberTKMKJH
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2025
Initial Report FDA Received Date03/19/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/20/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-