• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOLOGIC, INC. BIOZORB; MARKER, RADIOGRAPHIC, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC, INC. BIOZORB; MARKER, RADIOGRAPHIC, IMPLANTABLE Back to Search Results
Model Number F0304
Medical Device Problem Codes Detachment of Device or Device Component (2907); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/23/2017
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient received a lumpectomy procedure with placement of f0304 device on (b)(6) 2017.A new surgery required due to additional positive margins on (b)(6) 2017 and it was noted that the device was in 2 distinct pieces of roughly equal size in situ.The device was removed and a f0303 was used to replaced it.
 
Additional Manufacturer Narrative
Hologic is submitting this report in accordance with 21.Cfr part 803.The submission is based upon the currently available information.The submission of this report does not represent a confirmation of device malfunction involving the hologic products in the event described.A device history record (dhr) review was conducted for the reported lot/serial number.The device was released meeting all qa specifications.The device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOZORB
Common Device Name
MARKER, RADIOGRAPHIC, IMPLANTABLE
Manufacturer (Section D)
HOLOGIC, INC.
250 campus drive
marlborough MA 01752
Manufacturer (Section G)
HOLOGIC, INC.
250 campus drive
marlborough MA 01752
Manufacturer Contact
daniel guevara
coyol free zone building b24
alajuela
alajuela, 20102- CRI
CS   20102 CRI
MDR Report Key21645119
Report Number1222780-2025-01045
Device Sequence Number20772593
Product Code NEU
UDI-Device Identifier15420045514010
UDI-Public(01)15420045514010(17)180430(241)F0304(10)D1-151102
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K143484
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 02/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Expiration Date04/30/2018
Device Model NumberF0304
Device Catalogue NumberF0304
Device Lot NumberD1-151102
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/26/2017
Initial Report FDA Received Date03/19/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/02/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Hospitalization;
Patient SexFemale
-
-