| Model Number |
105-7100-060 |
| Medical Device Problem Codes |
Entrapment of Device (1212); Migration or Expulsion of Device (1395); Chemical Problem (2893); Unintended Movement (3026)
|
| Health Effect - Clinical Codes |
Paresis (1998); Perforation of Vessels (2135); Rupture (2208); Foreign Body In Patient (2687); Movement Disorder (4412); Decreased Level of Consciousness (4591)
|
| Date of Event |
03/11/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during the removal of the apollo catheter post embolization of an arteriovenous malformation, there was resistance and tension applied, leading to the rupture of the arteriovenous malformation (avm).The patient experienced motor and sensory loss to the right arm and leg, as detected by neuro monitoring.Vessel perforation occurred, and there was cast separation/movement during catheter removal.Onyx reflux was noted, and the catheter tip was difficult to remove, with the inner lining of the apollo catheter possibly left in the patient.A portion of onyx was displaced upon catheter removal.The devices were prepared according to the instructions for use (ifu).The catheter was flushed as per ifu.Surgery was planned for (b)(6) 2025.Ancillary devices: 5fr 45cm destination sheath, balt hybrid 12-14 guidewire, 038 127cm socrates as intermediate.
|
| |
|
Event or Problem Description
|
|
Additional information received reported raca not filling, potential ptfe liner still present.Laca collateralizing and patient is moving l upper extremity 3/5, no lower limb movement.There was no resistance upon injection of the liquid embolic agent into the apollo.During injection multiple pauses with total injection time of 9 minutes.The injection rate was followed as indicated in the instructions for use (ifu).Liquid embolic agent got embedded in an unintended location in the ra2.Additional intervention was post the embolization surgical resection of avm.The anatomical location of the apollo tip is the distal a3.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Additional information received reported the tip and what would seem to be the liner was detached.The inner liner is not referring to the detachable tip.2-4cm of reflux was noted.
|
| |
|