MAUDE Adverse Event Report: BIOMET 3I T3¿ PRO TAPERED IMPLANT 4/3MM (D) X 8.5MM (L); DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I T3¿ PRO TAPERED IMPLANT 4/3MM (D) X 8.5MM (L); DENTAL IMPLANT Back to Search Results
Catalog Number T3PT4385
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994); Osteolysis (2377)
Date of Event 01/30/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).H6: event problem codes: patient code: 1932.A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.H3 other text : summary investigation.
 
Event or Problem Description
Doctor reported that implant at tooth site 47 was removed due to peri-implantitis.Pain and inflammation were reported as a result of the event.Doctor also indicated a delay in the procedure of 8 weeks.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T3¿ PRO TAPERED IMPLANT 4/3MM (D) X 8.5MM (L)
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key21650620
Report Number0001038806-2025-00635
Device Sequence Number20703056
Product Code DZE
UDI-Device Identifier00844868046752
UDI-Public(01)00844868046752(17)290303(10)2120009738
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K213672
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 03/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue NumberT3PT4385
Device Lot Number2120009738
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/27/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date03/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/04/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
Patient SexFemale
Patient EthnicityNon Hispanic
-
-