Catalog Number 2112431
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/31/2025
Type of Reportable Event Malfunction
Event or Problem Description
As reported by the user facility: brief inquiry description: particulates found in final product.We have been noticing the occurrence of particulate matter being found at aql inspections at our caps lhv facility and had inquired about the availability of a filter to use for apex compounding.We were informed that a filter for apex is not available at this time.Would you be able to provide us with an official assessment that evaluates the possibility of using a filter or some kind of documentation as to why the use of a filter is not feasible at this time? the particles are found inside the finished final product.Recently, some of the particles that have been sent for identification testing have been identified as a synthetic latex rubber from the drug vial rubber stoppers.This would indicate the potential for coring of the rubber stoppers when the technician inserts the mini-spike into the vial.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow-up will be submitted when the investigation results become available.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).A review of the device history record (dhr) was performed for the reported lot number and no abnormalities or nonconformances were noted during the in process or final product inspection.No sample was provided for evaluation.Based on the data from the investigation we are unable to determine the root cause of the reported incident.The reported defect was unable to be confirmed.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINNACLE®
Common Device Name
CONTAINER, I.V.
Manufacturer (Section D)
B. BRAUN AVITUM ITALY S.P.A.
via xxv luglio 11
mirandola, 41037
IT  41037
Manufacturer (Section G)
B. BRAUN AVITUM ITALY S.P.A.
via xxv luglio 11
mirandola, 41037
IT   41037
Manufacturer Contact
jonathan severino
901 marcon blvd
allentown, PA 18109
4847197287
MDR Report Key21651099
Report Number9681240-2025-00036
Device Sequence Number9309035
Product Code KPE
UDI-Device Identifier04046964201141
UDI-Public(01)04046964201141(11)231101(17)261030(10)23L17
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K041415
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number2112431
Device Lot Number23L17
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/03/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/20/2025
Supplement Report FDA Received Date04/17/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-