| Model Number |
SUTURE ANCHOR, PEEK PUSHLOCK |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On 11th march 2025, it was reported by a sales rep via email that for 3 units of ar-1923ps, suture anchor, peek pushlock® 2.9 x 15.5 mm, on insertion of pushlock, the anchor crumbled up the inserter.This happened for all three units in the same batch.This was detected during use in a shoulder stabilization procedure on (b)(6) 2025.The procedure was completed successfully as they used a different batch of pushlock.13th march 2025: further information provided from the sales rep, after talking to the surgeon that one anchor broke while inserting into the drilled hole and pieces were retrieved by the surgeon.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: d2a.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to improper bone prep; misaligned insertion; prying/leveraging the device during insertion.
|
| |
|