• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. SWIVELOCK, SP PEEK 4.75MM; NON-DEGRADABLE FIXATN FASTENER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. SWIVELOCK, SP PEEK 4.75MM; NON-DEGRADABLE FIXATN FASTENER Back to Search Results
Model Number SWIVELOCK, SP PEEK 4.75MM
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/11/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a rotator cuff repair surgery the peek sp swivelock 4.75 tip broke in the patient.This happend twice with two different peek sp swivelocks 4.75mm.The broken off pieces have been retrieved from the patient.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with a different device ar-2324pslm.It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to improper bone prep; misaligned insertion; prying/leveraging the device during insertion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWIVELOCK, SP PEEK 4.75MM
Common Device Name
NON-DEGRADABLE FIXATN FASTENER
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key21653354
Report Number1220246-2025-01072
Device Sequence Number9246001
Product Code MAI
UDI-Device Identifier00888867384576
UDI-Public(01)00888867384576(11)240718(17)290630(10)15289810
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K222263
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSWIVELOCK, SP PEEK 4.75MM
Device Catalogue NumberAR-2324PSP
Device Lot Number15289810
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/11/2025
Supplement Date Received by Manufacturer09/03/2025
Initial Report FDA Received Date03/20/2025
Supplement Report FDA Received Date09/10/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/18/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-