SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD BENEVISION N17 PATIENT MONITOR Back to Search Results
Model Number BENEVISION N17 PATIENT MONITOR
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Output Problem (3005); No Visual Prompts/Feedback (4021)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/29/2025
Type of Reportable Event Death
Additional Manufacturer Narrative
A mindray field service representative collected the patient database, and the benevision n17 monitor logs for investigation.The log analysis shows the following: the n17 was set to standby mode at 15:28 on (b)(6) 2025 and exited standby mode at 7:39 on (b)(6) 2025.In standby mode, the monitor stops all parameter measurements and disables all alarm indications.In conclusion, there is no patient data for (b)(6) 2025, because the n17 was in standby mode.Benevision n17 performed according to specifications, no product malfunction occurred.
 
Event or Problem Description
The user reported being unable to see patient discharge data for (b)(6) 2025, on benevision n17 (serial number (b)(6).The user was able to see data on (b)(6) 2025, and (b)(6) 2025.The patient expired.
 
Additional Manufacturer Narrative
A mindray field service representative collected the patient database, and the benevision n17 monitor logs for investigation.The log analysis shows the following: the n17 was set to standby mode at 15:28 on (b)(6) 2025 and exited standby mode at 7:39 on (b)(6) 2025.In standby mode, the monitor stops all parameter measurements and disables all alarm indications.In conclusion, there is no patient data for (b)(6) 2025, because the n17 was in standby mode.Benevision n17 performed according to specifications, no product malfunction occurred.
 
Event or Problem Description
The user reported being unable to see patient discharge data for (b)(6) 2025, on benevision n17 (serial number (b)(6)).The user was able to see data on (b)(6) 2025, and (b)(6) 2025.The patient expired days after the (b)(6) 2025 (the exact date is unknown).
 
Additional Manufacturer Narrative
A mindray field service representative collected the patient database and the benevision n17 monitor logs for review.Mindray provided the benevision n17 keystroke activity for the required period.A mindray service representative, upon clarifying the request, returned to the site to collect logs from the anesthesia unit.However, since the attempt to collect data was on (b)(6) 2025, the data for the occurrence date ((b)(6) 2025) is no longer available on the a-series.No device malfunction was reported with the mindray devices.Units are monitored per specifications.
 
Event or Problem Description
During an investigation follow-up on april 7, 2025, the customer clarified the investigation request.No alleged malfunction was reported.The customer is requesting the keystroke for the a7 anesthesia machine (serial number (b)(6)) and the benevision n17 patient monitor (serial number (b)(6)).The request is for raw data from 09:00 am to 11:00 am on (b)(6) 2025.Some ventilation changes were made to the anesthesia machine at 10:28.Therefore, the customer requested to provide details on the keystroke activity on the anesthesia machine and bedside monitor 10 minutes before and 10 minutes after that time.The patient expired days after (b)(6) 2025 (the exact date is unknown).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BENEVISION N17 PATIENT MONITOR
Common Device Name
BENEVISION N17
Manufacturer (Section D)
SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD
1203 nanhuan avenue
guangming district
shenzhen, guangming 51810 6
CH  518106
Manufacturer Contact
sonia vale
800 macarthur blvd.
mahwah 07430
2013123286
MDR Report Key21654703
Report Number3009156722-2025-00005
Device Sequence Number7536539
Product Code MHX
UDI-Device Identifier06944904095578
UDI-Public06944904095578
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K213799
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberBENEVISION N17 PATIENT MONITOR
Device Catalogue Number6610E-PA00004
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2025
Supplement Date Received by Manufacturer04/15/2025
04/07/2025
Initial Report FDA Received Date03/20/2025
Supplement Report FDA Received Date04/18/2025
05/07/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient SexUnknown
-
-