Catalog Number 00585003012
Medical Device Problem Codes Fracture (1260); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Failure of Implant (1924); Joint Dislocation (2374)
Date of Event 02/26/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device was requested but not returned by the facility.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Additional associated products: 00585001396 polyethylene insert xt size c lot# 66309543.
 
Event or Problem Description
It was reported the patient was revised due to a fractured segmental hinge post and dislocation.Initial surgery date is unknown.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: b4, b5, d5, g3, g6, h1, h2, h3, h6, h10, h11.D10: additional associated products: 00585001396; polyethylene insert xt size c; lot# 66309543.00588000600; size 6 precoat cemented tibial component; lot# 65346357.00598801212; 12.7mm diameter 30mm l st stem length 75mm; lot# 65270143.00585004301; distal femoral xt component size c; lot# 66651240.00585004605; segment with male/female taper 50 mm length; lot# 65623591.00585004603; egment with male/female taper 30 mm length; lot# 66552395.00585204035; stem collar 35 mm o.D.; lot# 66134117.00585205015; fluted stem ext straight 15mm dia 130mm l; lot# 66054698.Visual examination of the provided images reveals that both devices, the hinge post, are fractured, various fragments are shown in the picture and the devices are disassembled.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The reported products were reviewed for compatibility with no issues noted.Medical records were not provided.This compliant can be confirmed based on the provided picture.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE WITH SEGMENTAL HINGE POST SIZE C 12MM
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key21655375
Report Number0001822565-2025-00697
Device Sequence Number9246049
Product Code KRO
UDI-Device Identifier00889024195745
UDI-Public(01)00889024195745(17)300301(10)65244819
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K150028
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number00585003012
Device Lot Number65244819
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/26/2025
Supplement Date Received by Manufacturer07/31/2025
Initial Report FDA Received Date03/20/2025
Supplement Report FDA Received Date08/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/18/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age25 YR
Patient SexMale
Patient Weight181 KG
-
-