• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. A.T.S. 3200TS TOURNIQUET W/HOSES (PERSONALIZED PRESSURE®); TOURNIQUET, PNEUMATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. A.T.S. 3200TS TOURNIQUET W/HOSES (PERSONALIZED PRESSURE®); TOURNIQUET, PNEUMATIC Back to Search Results
Catalog Number 60320010100
Medical Device Problem Codes Device Alarm System (1012); Measurement System Incompatibility (2982)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/27/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation into the reported event has been initiated under (b)(4).Once the investigation has been completed, a follow up report will be submitted with the investigation results and any actions taken by the manufacturer.
 
Event or Problem Description
It was reported that outside of surgery, the device flutters when using blue side, does not maintain consistent pressure and the unit was also beeping.There was no patient harm/injury or surgical delay as there was no patient involvement.The pressure was set up at: 250 mmhg.The pressure was not within the acceptable range for this unit as the actual pressure reading was between 243 -254mmhg.Due diligence is complete and there is no additional information available.
 
Additional Manufacturer Narrative
This complaint is recorded by zimmer biomet under (b)(4).Following sections were updated or corrected: b4, b5, d4, g1, g3, g6, h1, h2, h3, h4, h6 and h11.Review of the most recent repair record determined the software needed to be updated.The main pcb was replaced and resolved the reported issue.Dhr was reviewed and no discrepancies related to the reported event were found.The root cause of the reported issue is attributed to design issue.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
There is no additional information available regarding the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
A.T.S. 3200TS TOURNIQUET W/HOSES (PERSONALIZED PRESSURE®)
Common Device Name
TOURNIQUET, PNEUMATIC
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key21657331
Report Number0001526350-2025-00349
Device Sequence Number7480317
Product Code KCY
UDI-Device Identifier00889024626447
UDI-Public(01)00889024626447(11)230926(10)66318033
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number60320010100
Device Lot Number66318033
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/14/2025
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/27/2025
Supplement Date Received by Manufacturer08/04/2025
Initial Report FDA Received Date03/20/2025
Supplement Report FDA Received Date08/14/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/07/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-