| Catalog Number |
60320010100 |
| Medical Device Problem Codes |
Device Alarm System (1012); Measurement System Incompatibility (2982)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An investigation into the reported event has been initiated under (b)(4).Once the investigation has been completed, a follow up report will be submitted with the investigation results and any actions taken by the manufacturer.
|
| |
|
Event or Problem Description
|
|
It was reported that outside of surgery, the device flutters when using blue side, does not maintain consistent pressure and the unit was also beeping.There was no patient harm/injury or surgical delay as there was no patient involvement.The pressure was set up at: 250 mmhg.The pressure was not within the acceptable range for this unit as the actual pressure reading was between 243 -254mmhg.Due diligence is complete and there is no additional information available.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint is recorded by zimmer biomet under (b)(4).Following sections were updated or corrected: b4, b5, d4, g1, g3, g6, h1, h2, h3, h4, h6 and h11.Review of the most recent repair record determined the software needed to be updated.The main pcb was replaced and resolved the reported issue.Dhr was reviewed and no discrepancies related to the reported event were found.The root cause of the reported issue is attributed to design issue.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
There is no additional information available regarding the event.
|
| |
|