| Catalog Number |
RSF-295+ |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Burning Sensation (2146); Breast Discomfort/Pain (4504); Insufficient Information (4580)
|
| Date of Event |
01/25/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
As stated in the literature, ¿a number of risk factors for rupture have been identified; the most common cause is surgical instrument damage.¿ (handel, garcía and winxtrom, 2013.Breast implant rupture: causes, incidence, clinical impact, and management.Plast.Reconstr.Surg.132: 1128.) adittionally, rupture is a common and known reason for complaints, and it is clearly characterized in the product dfu included with the implant, as stated above.Additionally, no cases of rupture because of product failure have been reported to establishment labs ever.Per our directions for use, each patient must receive the establishment labs s.A.¿motiva implants®: information for the patient¿ during her surgical consultation, it is the surgeons´ responsibility to ensure that the patient completely understands the information regarding the risks, benefits, and recommendations associated with silicone gel-filled breast implant surgery, as well as the complications typical of any type of surgery.This document is available in motiva® website: ifu.Motiva.Health.The instructions for use in the directions for use were reviewed, to determinate if there are indications that ensure the prevention and good handling of the product.For this event, it is considered that the information is suitable for the section of surgical precautions as follows: ¿breast implants can potentially remain intact for decades in the body, but all such devices will fail at some point.Breast implants rupture when the shell develops a tear or hole.Rupture can occur at any time after implantation, but it is most likely to occur; the longer the implant is in place.The following may cause implants to rupture: damage by surgical instruments, implant stress, and weakening during implantation, age and design of the implant, submuscular rather than sub-glandular location, the occurrence of post-operatory hematomas or seromas, folding or wrinkling of the implant shell, excessive force to the chest, trauma, compression during mammographic imaging, and severe capsular contracture¿.Establishment labs requested the unit to confirm the event and to perform the corresponding testing (including but not limited to: revision and characterization under the microscope of the rupture section, mechanical testing according to approved standards, etc.).It was informed that the patient/doctor has the unit, and it is pending to confirm if it can be returned.Breast implants are not considered lifetime devices, and implant rupture is a risk associated with breast surgery, which can occur at any time.In addition, a review of the device history record was conducted, and it was concluded that there were no deviations in the manufacturing process of this device that would have caused or contibuted to this incident.When the investigation process is completed, the applicable findings will be included in a supplemental medwatch.At establishment labs, we are committed to patient safety and continually evaluate the performance of our devices through postmarket surveillance of reported complaints & adverse events.
|
| |
|
Event or Problem Description
|
|
(b)(6).It was reported that, a patient who was implanted in (b)(6) 2023, presented a high level of pain and burning sensation due to a device rupture on the left side implant.Both implants were replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Quality department (pms) has received a complaint regarding a device from establishment labs, additional information provided in the "details" section.General conclusion: a relationship between the alleged event and the device involved: no event analysis: the corresponding test were not performed due to the unit involved was not returned for analysis.Additionally, rupture is a common and known reason for complaints, and it is clearly characterized in the product dfu included with the implant, as stated above.Breast implants are not considered lifetime devices, and implant rupture is a risk associated with breast surgery, which can occur at any time.Dhr review: a complete review of the dhr for lot involved was carried out, no deviation was found in the manufacturing process.Additionally, there were no indications of abnormalities in raw materials or manufacturing processes which may have affected this particular batch.Refers to phr section dfu review: the instructions for use in the directions for use were reviewed, to determinate if there are indications that ensure the prevention and good handling of the product.For this event, it is considered that the information is suitable for the section of surgical precautions as follows: rupture breast implants rupture when the shell develops a tear or hole.Rupture can occur at any time after implantation but is more likely to occur the longer the implant is in place.The following may cause implants to rupture: damage by surgical instruments, implant stress and weakening during implantation, implant age and design, submuscular rather than subglandular location, the occurrence of post-operatory hematomas or seromas, folding or wrinkling of the implant shell, excessive force to the chest, trauma, compression during mammographic imaging, and severe capsular contracture.Per our directions for use, each patient must receive the establishment labs s.A.¿motiva implants®: information for the patient¿ during her surgical consultation, it is the surgeons´ responsibility to ensure that the patient completely understands the information regarding the risks, benefits, and recommendations associated with silicone gel-filled breast implant surgery, as well as the complications typical of any type of surgery.This document is available in motiva® website: ifu.Motiva.Health.As stated in the literature: ¿a number of risk factors for rupture have been identified; the most common cause is surgical instrument damage.¿ (handel, garcía and winxtrom, 2013.Breast implant rupture: causes, incidence, clinical impact, and management.Plast.Reconstr.Surg.132: 1128.) trend review: establishment labs will continue to monitor for similar complaints/trends.As part of the post market surveillance process, monitoring of the primary endpoints reported is performed to determine unfavorable trends.Per our current complaint data report, no unfavorable trends were detected on this product.No further actions are required.
|
| |
|
Additional Manufacturer Narrative
|
|
1.Overview the quality department (pms) received a complaint involving a motiva® device.Further details are provided in the cq ¿details¿ section.2.General conclusion device involvement: a causal relationship between the reported event and the device has been established.Event characterization: the event was classified as a device rupture.3.Technical investigation summary laboratory testing laboratory evaluation was conducted in accordance with wi-001048 "pms laboratory testing units." key findings include: visual inspection: macroscopic examination of the returned device revealed clear evidence of shell rupture.Microscopic analysis: under magnification, the shell exhibited trace marks indicative of interaction with a sharp instrument.The morphology of these marks closely matched those reproduced under controlled laboratory conditions simulating instrument-related damage.These findings differ significantly from patterns associated with spontaneous rupture (e.G., fatigue or shell degradation).Shell compliance testing: mechanical and material integrity testing confirmed that the implant shell met all applicable international specification standards.Reference: for-001037 ¿ complaint investigation unit testing results.Root cause analysis based on the testing results, the most probable root cause of the iatrogenic rupture is a damage from a sharp surgical instrument.Such damage may have compromised the shell¿s integrity, predisposing it to failure during clinical procedure.4.Dfu and labeling evaluation the alleged event is a known, well-documented complication associated with silicone breast implants.It is clearly addressed in the product¿s directions for use (dfu), which states: ¿breast implants can rupture when the shell develops a tear or hole.Rupture can occur at any time after implantation, but it is more likely to occur over time.Silent rupture is common and often asymptomatic; therefore, patients should undergo regular mri screenings starting at 3 years post-op, then every 2 years.¿ the dfu also emphasizes the responsibility of the surgeon to fully inform the patient of potential risks and complications during the pre-operative consultation.Patients are provided with the document ¿motiva implants®: information for the patient,¿ accessible at ifu.Motiva.Health.Breast implants are not considered lifetime devices, and implant rupture is a risk associated with breast surgery, which can occur at any time.Literature reference: handel, n., garcía, m.E., & wixtrom, r.(2013).Breast implant rupture: causes, incidence, clinical impact, and management.Plast.Reconstr.Surg.132(5), 1128."a number of risk factors for rupture have been identified; the most common cause is surgical instrument damage." 5.Device history record (dhr) review a comprehensive review of the device history record (dhr) for the affected lot was completed.No deviations, non-conformances, or anomalies were identified during manufacturing.All raw materials and process parameters conformed to specified quality requirements.6.Trend and signal monitoring the event has been evaluated as part of ongoing trend analysis activities within the pms program.Per the current complaint data report: no adverse or unexpected trends related to the alleged event have been identified.No further corrective or preventive actions are required at this time.Establishment labs will continue to monitor for similar events as part of routine pms surveillance.
|
| |
|
Additional Manufacturer Narrative
|
|
During our internal quality review of the submitted mdr, we identified an unintentional misclassification of the event.The report was originally submitted under the ¿malfunction¿ category.However, according to fda¿s event classification criteria, the event should have been categorized as serious injury, as the patient experienced a condition that required medical intervention to prevent permanent impairment.This misclassification was due to an administrative oversight during the initial assessment of the complaint information.Upon further evaluation of the clinical details received, we confirmed that the reported harm meets the definition of a serious injury per 21 cfr 803.3.We have taken corrective measures to update the event classification accordingly and ensure that the mdr accurately reflects the nature of the incident.
|
| |
|