| Model Number |
0015807 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/20/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported the stretcher is intermittently lowering abruptly when pressing the down button which results in startling the patients and crew.No injuries have been reported and no specific patient incident was cited.
|
| |
|
Additional Manufacturer Narrative
|
|
The subject stretcher was returned to manufacturer for a visual and functional evaluation.After multiple tests, the alleged intermittent lowering was able to be confirmed.The actuator was identified as the contributing factor and was replaced.
|
| |
|
Search Alerts/Recalls
|