| Model Number |
1012273-20 |
| Medical Device Problem Codes |
Material Rupture (1546); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint history identified no other similar complaints from this lot.Based on the information received, the investigation determined that the reported complaints appear to be related to operational context.There was no leak noted during preparation, which suggests a product quality issue did not contribute to the reported difficulties.It was reported by the account that the balloon dilatation catheter (bdc) interacted with the moderately calcified and heavily tortuous anatomy resulting in the reported difficulty advancing the device and subsequently the balloon ruptured during inflation.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a lesion in the left circumflex (lcx) artery with moderate calcification and heavy tortuosity.For pre-dilatation, the 2.75x20mm trek balloon dilatation catheter (bdc) was advanced to the target lesion with resistance due to the anatomy and the balloon was inflated; however, the balloon ruptured during the first inflation at 6 atmospheres (atm).Therefore, the bdc was removed from the patient.A non-abbott balloon was used to continue the procedure.There was no adverse patient effect and no clinically significant delay reported in the procedure.No additional information was provided.
|
| |
|
Search Alerts/Recalls
|