ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752217
Medical Device Problem Code Suction Problem (2170)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/20/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A health care professional reported pak was defective because there was no suction during (at the start) of cataract surgery (with intraocular lens implant).The surgery was completed with another pak.There was no patient impact.
 
Additional Manufacturer Narrative
A used fluidics management system (fms) was visually inspected and was tested using a console.The product could be recognized, and the service data could be retrieved from the console.However, the product failed to prime and generated a system message code.The cassette and manifolds were purged; no occlusion was observed.The product was re-tested and the same system message code occurred.The drip chamber was replaced with a lab stock drip chamber.The product passed priming.A submerge fluid test for the returned drip chamber was performed by purging air from the syringe.No air came from the filter of the drip chamber spike.The investigation conducted a non-conformance review of the reported lot number.No deviations that could have contributed to this event were identified and all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information and materials received, the investigation concluded that the root cause of the reported event is external supplier related, caused by an error during the suppliers manufacturing process.The supplier has been notified of this complaint and will take appropriate actions as necessary.All lots are verified that all required inspections have been performed and all acceptance criteria are met.Complaint data for all products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION SURGICAL PROCEDURE PAK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key21661222
Report Number1644019-2025-00763
Device Sequence Number18330816
Product Code LRO
UDI-Device Identifier00380657522170
UDI-Public(01)00380657522170(17)260731(11)240825(10)170YHX
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K880961
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2026
Device Catalogue Number8065752217
Device Lot Number170YHX
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 02/24/2025
Supplement Date Received by Manufacturer05/23/2025
Initial Report FDA Received Date03/21/2025
Supplement Report FDA Received Date06/13/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/25/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BSS STERILE IRRIGATING SOLUTION
Patient SexUnknown
-
-