| Model Number |
ACT200 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/20/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Medtronic received information that at an unspecified time, this act plus instrument had an average activated clotting time measurement that was 100 below what other devices measured.The use of the instrument was unknown.There was no adverse patient effect reported with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic received additional information that the customer could not provide additional details on the case.It was stated that they have only registered consistently that the activated clotting time (act) measurements are often quite a lot lower than expected based on the given heparin dose.They have tested with act instruments from other sections at the cardiac department and have found that they have measured higher act at the same time as this act instrument has measured up to 100 in act difference (lower).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evlauation: the reported issue that the instrument had an average activated clotting time measurement that was 100 below what other devices measured was verified during service.Service technician observed a loose contact on the heat block grounding.The lift bar height was out of specifications.Service technician found errors in statistic log file: 010, 012, 064 and 069 and observed that the disk drive could not read disks.The issue will be resolved by fitting the heat block grounding, adjusting the lift bar height from 0.0620-0.0640 inch to 0.0915-0.0935 inch, repairing the disk drive and replacing the overlay, assy switch, and keypad overlay.Service technician will perform the cartridge test hr-nm 100-96.4 sec, (range 080-120 sec) and hr-ab 279-285 sec, (range 267-559 sec) and clear statistic log file.Preventative maintenance will be performed per specifications.D.4.Udi updated medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|