| Catalog Number |
09556508001 |
| Medical Device Problem Code |
False Negative Result (1225)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/24/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The eplex base unit serial numbers were (b)(6).Sampler material from the patient was requested for further investigation.The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
There was an allegation of false negative staphylococcus aureus results for two samples from the same patient using the eplex blood culture identification panel - gram positive (bcid-gp) panel results with the eplex base units.The samples were from the same patient but different bottle sets.The tests were performed on (b)(6) 2025.Sample 1 was tested on analyzer eplex serial number (b)(6) multiple times and then tested on eplex serial number (b)(6).Sample 2 was only tested on eplex serial number (b)(6).All of the results from the eplex analyzers were staphylococcus species.The culture results were staphylococcus aureus (mssa).
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation did not identify a product problem.The event was consistent with sample-related factors, such as a lower than intended target concentration within the sample volume analyzed by the assay, or with variations within the intended test target.There is a risk of false-negative results due to the innate nature of microbes.
|
| |
|
Additional Manufacturer Narrative
|
|
Medwatch field d4 lot no was updated.The two patient samples were received for further investigation.Both samples were tested, and staphylococcus and staphylococcus aureus were detected.The investigation did not identify a product problem.The cause of the event could not be determined, but the event was consistent with sample-related issues.It was possible that a lower than intended target concentration was within the sample volume analyzed by the assay.
|
| |
|
Search Alerts/Recalls
|