Model Number FG000016-05
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code Elevated ketones/Diabetic Ketoacidosis (2364)
Date of Event 02/04/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H11: since no lot number is available, a detailed investigation, tests on reference samples or batch review cannot be conducted.Therefore, this complaint will be closed, this issue will be monitored through pms product trends and malfunction.If any trends picked up, this will flag on the trips and alerts according to the omqr procedure.
 
Event or Problem Description
Reference number (b)(4).Event occurred in the united states.It was reported that the patient faced diabetic ketoacidosis due to bent cannula event went to the emergency room (er) and eventually got hospitalized and was admitted into the intensive care unit (icu) on (b)(6) 2025.The blood glucose level was 386 mg/dl at the time of event and the patient got treated with multiple daily injection (mdi).The patient was tested with ketones and the results were high.Company do not see bent/kinking as being related to human factors, but rather as a training issue including correct choices of insertion sites and infusion sets and cannula length.Furthermore, the soft cannula is a flexible material that during use and upon removal can bend slightly.No further information available.
 
Additional Manufacturer Narrative
Unomedical hereby submits this supplemental report as part of its complaint remediation activities conducted under a capa/fda action plan.This submission includes a retrospective reassessment of previously evaluated complaints.Unomedical is providing this supplemental information in accordance with the reporting requirements set forth in 21 cfr part 803.Any fields left blank indicate that the information is unknown, unavailable, or remains unchanged.Additional information - this mdr is being submitted to include the below: h6: investigation results under type of investigation, investigation findings, investigation conclusions.H11: investigation summary.Complaint investigation results.A complaint investigation was conducted based on the event description and the assigned malfunction code soft cannula bent/kinked/crimped after removal -blockage.Additional information was requested to support the investigation; however, a lot number information was not provided.No device, device components, or other visual or physical evidence was made available for evaluation.Consequently, visual inspection, retain-sample testing, or the assessment of potential product performance issues, component integrity, or manufacturing defects could not be performed.In response to the complaint and due to the absence of a specific lot number, a high level investigation was conducted for the inset product family and the assigned malfunction.The investigation encompassed an · corrective and preventive action (eqms) search, assessment of complaint trends, and the review of risk management files.No systemic issues were identified.Please refer to the attached memo for full details: inset icd-pmc05 lopi cannula.Enclosure 1: eqms search results.Enclosure 2: maintenance results.Enclosure 3: raw data for complaint trending.Corrective and preventive action (capa)determination.Based on the investigation, no further action is required.The record will be closed and monitored through tracking and trending per work instruction (wi) (monthly trips and alerts).Complaint unconfirmed: following investigation, the reported failure could not be confirmed for this complaint.Complaint investigation - conclusion.Due to the absence of a lot number and returned product, the investigation was limited to a high level review of the inset product family, including an eqms search, complaint trending, and review of applicable risk management documentation.No evidence of a systemic issue was identified.No further action is required at this time, and the record will be closed with continued monitoring through routine tracking and trending.The record may be reassessed if additional information becomes available.
 
Event or Problem Description
To date no additional patient or event details have been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSET
Common Device Name
UNO INSET I 80/6 GREY BETA 10PK BETA
Manufacturer (Section D)
UNOMEDICAL UM-D
ave. fomento industrial l9 m3
parque industrial del norte
reynosa , tamps, 88736
MX  88736
Manufacturer Contact
aaholmvej 1-3
osted
lejre, 4320
MDR Report Key21662482
Report Number3003442380-2025-04339
Device Sequence Number15688519
Product Code FPA
UDI-Device Identifier05705244026254
UDI-Public05705244026254
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K032854
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 01/26/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberFG000016-05
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2025
Supplement Date Received by Manufacturer01/26/2026
Initial Report FDA Received Date03/21/2025
Supplement Report FDA Received Date02/04/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-