• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. THREE PEG PATELLA 29MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. THREE PEG PATELLA 29MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 200-02-29
Medical Device Problem Codes Unstable (1667); Naturally Worn (2988); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Pain (1994); Synovitis (2094); Discomfort (2330); Osteolysis (2377); Ambulation or Postural Difficulties (2544); Joint Laxity (4526); Swelling/ Edema (4577); Joint effusion (4849)
Date of Event 08/18/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported via legal documentation that approximately 171 months after a left total knee replacement procedure, the patient underwent a revision procedure to address loosening, prothesis wear, instability, ambulation difficulties, discomfort, joint effusion, joint laxity, osteolysis, pain, swelling/edema and synovitis.Revision surgery operative notes were provided.Patient left the operating room in stable condition.No further issues or complications were reported.No additional information is available.
 
Additional Manufacturer Narrative
D10: (b)(6) 204-04-22 - trapezoid tibial tray sz 1f/2t, 2f/2t, (b)(6) 204-22-11 - ps tibial inserts sz 2, 11mm, (b)(6) 234-02-02 - optetrak asy, ps cemented femoral, sz 2.Multiple mdr reports were filed for this event, please see associated reports: 1038671-2023-00152.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, instability, and patellar loosening or due to inclusion of the polyethylene in the packaging recall.Additionally, these devices appear to have been implanted for over ten years prior to the reported event.Potential contributions of manufacturing, user, and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and product information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THREE PEG PATELLA 29MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
MDR Report Key21662653
Report Number1038671-2025-01634
Device Sequence Number8431534
Product Code JWH
UDI-Device Identifier10885862039583
UDI-Public10885862039583
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K932690
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/20/2012
Device Catalogue Number200-02-29
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/20/2023
Initial Report FDA Received Date03/21/2025
Date Device Manufactured08/23/2007
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age53 YR
Patient SexFemale
-
-