| Catalog Number |
8065752917 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/20/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A physician reported that the actuation failure of the vent valve occurred during cataract surgery with intraocular lens implantation.The procedure was completed by replacing the product with new one.There was no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
The fluidics management system (fms) in the tray was visually inspected.The drain bag was detached from the drain bag tape on the cover due to saturation.The drain bag tape was adhered on the cassette properly.Both irrigation and aspiration luers were intact.Flare shape was observed in the blue socket fitting at the aspiration tubing insertion end.The product was tested using a calibrated console with the current software.The product could be recognized by the console.The product primed and tuned with the handpiece and the 0.9-millimeter aspiration bypass system (abs) tip and infusion sleeve from the lab stock, successfully and the service data could be retrieved.No system message code displayed on console screen.The product functional testing was all within specification.The investigation conducted a non-conformance review of the reported lot number.No deviations were identified, and all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information and materials received, the investigation was unable to identify the root cause or origin of the reported event.The product functioned per specifications.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint.No further investigative or manufacturing actions are warranted at this time due to the product meeting specifications.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|