MAUDE Adverse Event Report: ARTHREX, INC. ANGLED REAMER SLEEVE, 10°; MANUAL INSTR, GENERAL SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. ANGLED REAMER SLEEVE, 10°; MANUAL INSTR, GENERAL SURGICAL Back to Search Results
Model Number ANGLED REAMER SLEEVE, 10°
Medical Device Problem Code Device Damaged by Another Device (2915)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/06/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On 03/07/2025, a sales representative reported via (b)(4) that an ar-9676 angle reamer drive shaft spun and stripped inside the ar-9597-10 angle reamer sleeve 10 degrees while reaming the glenoid.They had to leave the glenoid as it was.The reversal was completed in full.It did add 5 minutes to the case.This was discovered during a revers procedure, with no reported patient harm.Additional information was received on 03/13/2025: they had to leave the glenoid as it was reamed and could not use another option.There was no report of adverse effects on or after the surgery for the patient.This occurred during an rtsa procedure on (b)(6) 2025.
 
Additional Manufacturer Narrative
Additional information: the complaint allegation is confirmed.One unpackaged ar-9597-10, serial/batch number (b)(6), was received for investigation.The visual evaluation noted abrasion marks along the inside of the shaft and heavy circumferential grind marks on the proximal end of the device.Functional testing was not performed due to the damage to the device.The most likely cause for the reported failure can be attributed to misuse due to a force perpendicular to the drive shaft being applied while the device is spinning.This applied force creates an excess of friction and heat that ultimately has the potential to cause the device to cease functioning as intended.Application of this force perpendicular to the drive shaft is not needed for the device to function, nor is it recommended that the user do this during normal clinical use.When used normally, this issue is unlikely to present itself, which suggests that the handling of the device greatly impacts whether or not it will seize.For this reason, the direct cause is considered to be a misuse of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGLED REAMER SLEEVE, 10°
Common Device Name
MANUAL INSTR, GENERAL SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key21664380
Report Number1220246-2025-01103
Device Sequence Number18070886
Product Code LXH
UDI-Device Identifier00888867343436
UDI-Public00888867343436
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 07/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberANGLED REAMER SLEEVE, 10°
Device Catalogue NumberAR-9597-10
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/07/2025
Supplement Date Received by Manufacturer06/09/2025
Initial Report FDA Received Date03/21/2025
Supplement Report FDA Received Date07/02/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-