EXACTECH, INC. THREE PEG PATELLA 29MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
200-02-29 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/26/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(d10) concomitant devices: 02-010-03-0230 - logic cr femoral cem, left, sz 3: (b)(6).02-012-45-3040 - lgc tibial fit tray cem sz 3f / 4t: (b)(6).02-012-49-3011 - logic cr tib insert slope++, sz 3, 11mm: (b)(6).The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
As reported, approximately six years post initial left side tka, the patient presented back and was scheduled for a revision of the total knee possible poly swap due to poly wear, failure, and recall.A tibial insert poly swap and a patella implant revision were performed.There was a 30-45 minutes surgical delay/prolongation; however, patient did not experience any adverse events as a result of surgical prolongation.Patient was last known to be in stable condition following the event.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional and/or corrected information.The reason for the revision reported was likely the result of prosthesis wear and inclusion of the polyethylene in the packaging recall.Additionally, a contributing factor to the reported wear may have been the result of being packaged in a non-conforming vacuum bag for more than five years.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and relevant clinical information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|