Product complaint # (b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 device not returned.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: please provide the patient's demographic information including age, gender, weight, bmi at the time of index procedure 27y/o female, 77.27 kg, bmi 30.1.Date and name of index surgical procedure? (b)(6) 2025, r ltcs.The diagnosis and indication for the index surgical procedure? previous ltcs.What was the initial approach for the index surgical procedure? (open, laparoscopic or other)? open.On what tissue was the suture used/location of suture placement? fascia.What tissue dehisced? fascia.What was the tissue condition (normal, thin, calcified, fragile, diseased)? how was the suture placed (interrupted or continuous)? continuous.How was the suture tied (square knot or multiple knots one end)? onset date/time of dehiscence? (# post op days) 2/11/25 at 1320.What symptoms did the patient experience following the index surgical procedure? bleeding from incision site.Onset date? 2/11/25.How was the dehiscence managed? surgical repair.It was stated that the patient ¿returned for a wound dehiscence¿.Yes, please provide further clarification and details on this statement.Fascial dehiscence after c/s.Please describe any surgical intervention required for the wound dehiscence including date and findings.Repair of fascial dehiscence.Please describe the appearance of the suture during the second procedure.Pds snapped in the midline of the incision, omentum extruded in the subcutaneous space.Is it known how the wound dehisced? suture popped.Did the suture pull out of the tissue? did the suture break? yes, if so, where was the break noted (termination, middle, end)? middle.Were there any precipitating patient stress factors for the sutures untying, breakage or pulling out of the tissue? did the operating surgeon observe any suture deficiency or anomaly before, during, after the suture placement and during any re-operation? no.Other relevant patient history/concomitant medications? what is the physician¿s opinion as to the etiology of or contributing factors to this event? what is the patient's current status? stable.Surgeon¿s name? dr.(b)(6).
|
|
It was reported that a patient underwent an ltcs on (b)(6) 2025 and suture was used.On the same day, (b)(6) 2025, the patient had bleeding from incision site.The suture snapped in the midline of the incision, omentum extruded in the subcutaneous space.Patient had a fascial dehiscence and underwent a repair of the fascial dehiscence.Additional information was requested.
|
|
Product complaint # (b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected d3 manufacturer email.Corrected h6 type of investigation.A manufacturing record evaluation was performed for the finished device lot number, and no related non-conformances were identified.
|