MAUDE Adverse Event Report: ETHICON INC. PDS PLUS SUTURE 36"(90CM) 0 VIO; SUTURE, SURGICAL, ABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. PDS PLUS SUTURE 36"(90CM) 0 VIO; SUTURE, SURGICAL, ABSORBABLE Back to Search Results
Catalog Number PDP370T
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Wound Dehiscence (1154); Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 02/11/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H6 component code: g07002 device not returned.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: please provide the patient's demographic information including age, gender, weight, bmi at the time of index procedure 27y/o female, 77.27 kg, bmi 30.1.Date and name of index surgical procedure? (b)(6) 2025, r ltcs.The diagnosis and indication for the index surgical procedure? previous ltcs.What was the initial approach for the index surgical procedure? (open, laparoscopic or other)? open.On what tissue was the suture used/location of suture placement? fascia.What tissue dehisced? fascia.What was the tissue condition (normal, thin, calcified, fragile, diseased)? how was the suture placed (interrupted or continuous)? continuous.How was the suture tied (square knot or multiple knots one end)? onset date/time of dehiscence? (# post op days) 2/11/25 at 1320.What symptoms did the patient experience following the index surgical procedure? bleeding from incision site.Onset date? 2/11/25.How was the dehiscence managed? surgical repair.It was stated that the patient ¿returned for a wound dehiscence¿.Yes, please provide further clarification and details on this statement.Fascial dehiscence after c/s.Please describe any surgical intervention required for the wound dehiscence including date and findings.Repair of fascial dehiscence.Please describe the appearance of the suture during the second procedure.Pds snapped in the midline of the incision, omentum extruded in the subcutaneous space.Is it known how the wound dehisced? suture popped.Did the suture pull out of the tissue? did the suture break? yes, if so, where was the break noted (termination, middle, end)? middle.Were there any precipitating patient stress factors for the sutures untying, breakage or pulling out of the tissue? did the operating surgeon observe any suture deficiency or anomaly before, during, after the suture placement and during any re-operation? no.Other relevant patient history/concomitant medications? what is the physician¿s opinion as to the etiology of or contributing factors to this event? what is the patient's current status? stable.Surgeon¿s name? dr.(b)(6).
 
Event or Problem Description
It was reported that a patient underwent an ltcs on (b)(6) 2025 and suture was used.On the same day, (b)(6) 2025, the patient had bleeding from incision site.The suture snapped in the midline of the incision, omentum extruded in the subcutaneous space.Patient had a fascial dehiscence and underwent a repair of the fascial dehiscence.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected d3 manufacturer email.Corrected h6 type of investigation.A manufacturing record evaluation was performed for the finished device lot number, and no related non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PDS PLUS SUTURE 36"(90CM) 0 VIO
Common Device Name
SUTURE, SURGICAL, ABSORBABLE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
9083863534
MDR Report Key21664945
Report Number2210968-2025-03054
Device Sequence Number7614564
Product Code NEW
UDI-Device Identifier10705031048195
UDI-Public(01)10705031048195(17)250630(10)THMAXQ
Combination Product (Y/N)Y
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K061037
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2025
Device Catalogue NumberPDP370T
Device Lot NumberTHMAXQ
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/26/2025
Supplement Date Received by Manufacturer04/09/2025
Initial Report FDA Received Date03/21/2025
Supplement Report FDA Received Date04/17/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/06/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age27 YR
Patient SexFemale
Patient Weight77 KG
-
-