|
Reportable based on device analysis completed on february 26, 2025.It was reported that the device failed to inflate.The target lesion, which was 50% stenosed, was located in a moderately tortuous and mildly calcified vessel below the knee.A 3.0 x 60 x 150 mm sterling balloon catheter was advanced to the site for dilation.During the procedure, an attempt was made to inflate the balloon; however, the pressure continuously dropped, and the balloon did not inflate as expected.The device was subsequently removed, and further attempts to inflate the balloon outside the body confirmed the presence of a leak.The procedure was successfully completed using an alternative device, and no patient complications were reported.Analysis of the returned device revealed the presence of a longitudinal tear.
|