MAQUET CARDIOVASCULAR LLC VASOVIEW 6 PRO; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number VH-2400
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/27/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has not yet been returned to maquet cardiac surgery for evaluation.We are following up with the customer for the return of the device.A supplemental report will be submitted if the device is received.
 
Event or Problem Description
The hospital reported that during an endoscopic vein harvesting procedure, vasoview 6 pro c-ring was detached from one of the arms.They opened a new device.There was a 5 minute delay.No harm to the patient.No parts left behind.
 
Additional Manufacturer Narrative
Tw # (b)(4).Updated sections: b4, d9, g3, g6, h2, h3, h6, h11.The lot # 3000430443 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASOVIEW 6 PRO
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ 07470
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 07470
MDR Report Key21666805
Report Number2242352-2025-0000337
Device Sequence Number7756725
Product Code GEI
UDI-Device Identifier00607567700239
UDI-Public(01)00607567700239(10)3000430443(17)261104
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K091733
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVH-2400
Device Catalogue NumberVH-2400
Device Lot Number3000430443
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/27/2025
Supplement Date Received by Manufacturer06/20/2025
Initial Report FDA Received Date03/21/2025
Supplement Report FDA Received Date06/23/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/04/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NO
Patient SexUnknown
-
-