• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FERNO-WASHINGTON, INC. TRANSCEND W/TRX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FERNO-WASHINGTON, INC. TRANSCEND W/TRX Back to Search Results
Model Number 0731402
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/05/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported one of the front transport wheels on the chair became loose and detached from the frame during a patient transport.No injuries or adverse effects were reported as a result.The chair was evaluated by an authorized field technician and the issue was confirmed.It was noted the shank pin on the wheel was bent.The caster was replaced and the chair was returned to service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANSCEND W/TRX
Common Device Name
TRANSCEND W/TRX
Manufacturer (Section D)
FERNO-WASHINGTON, INC.
70 wiel way
wilmington OH 45177
Manufacturer (Section G)
FERNO-WASHINGTON, INC.
70 wiel way
wilmington OH 45177
Manufacturer Contact
dawn greene
70 wiel way
wilmington, OH 45177
9372832900
MDR Report Key21667701
Report Number1523574-2025-00044
Device Sequence Number12806815
Product Code FPP
UDI-Device Identifier00190790004057
UDI-Public(01)00190790004057(11)230320
Combination Product (Y/N)N
Initial Reporter CountryCA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 03/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0731402
Device Catalogue Number0731402
Device Lot NumberM511880
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2025
Initial Report FDA Received Date03/21/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/20/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-