BECTON DICKINSON BD CATHENA 20GX1.00IN STRAIGHT BC; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 386803
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/05/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If a device evaluation and/or device history review is completed, a supplemental report will be filed.B3.The date received by manufacturer has been used for this field.
 
Event or Problem Description
It was reported that bd cathena 20gx1.00in straight bc leaked it was reported by customer that they encounter occasional problem with the product 386803 and batch 4148272.Verbatim: rcc received a complaint via email.Email(s) attached.A requester informs me of an occasional problem with the product 386803 having the batch # of 4148272.
 
Additional Manufacturer Narrative
No photos or samples were received by our quality team for evaluation therefore the failure mode could not be verified.A review of the internal manufacturing device records and raw material history files for the reported lot numbers was performed and no recorded quality problems or rejections to this incident were found.Corrective actions have been initiated, and a manufacturing root cause has been identified for this failure mode.Actions have been made to address the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD CATHENA 20GX1.00IN STRAIGHT BC
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore NJ 07417
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key21668135
Report Number2243072-2025-00355
Device Sequence Number19933981
Product Code FOZ
UDI-Device Identifier00382903868032
UDI-Public(01)00382903868032(11)240625(17)270531(10)4148272
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K172506
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number386803
Device Lot Number4148272
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/03/2025
Supplement Date Received by Manufacturer04/28/2025
Initial Report FDA Received Date03/21/2025
Supplement Report FDA Received Date05/14/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/25/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-