|
Pneumothorax is the most common side effect associated with the zephyr valve treatment (criner et al.A multicenter randomized controlled trial of zephyr endobronchial valve treatment in heterogeneous emphysema (liberate).Am j respir crit care med.2018; 198 (9): 1151-1164).Targeted lobar deflation likely causes inflation of the ipsilateral lobe, which can result in a tear of the already compromised parenchymal tissue of the emphysematous ipsilateral lobe, resulting in a pneumothorax (criner et al.A multicenter randomized controlled trial of zephyr endobronchial valve treatment in heterogeneous emphysema (liberate).Am j respir crit care med.2018; 198 (9): 1151-1164).In the liberate study (ide clinical study used to support pma p180002's approval), 26.6% of the zephyr valve subjects experienced a pneumothorax in the treatment period ([less than or equal to 45 days).These were managed using standard of care procedures as per previously published guidelines (valipour, arschang, et al.Respiration 87.6 (2014): 513-521).In 17.4% of the events, the pneumothorax resolved without any additional intervention with subjects under careful observation.In over half the events (56.5%), the pneumothorax was managed with a chest-tube only.An additional 13% of the events were managed with a chest-tube and the temporary removal of one or more valves, while another 13% of the events were managed with a chest-tube and removal of all the implanted valves.Upon successful resolution of the pneumothorax, removed valves can be replaced.Patients that experienced a pneumothorax experienced clinical benefits of the zephyr valve treatment that were similar to the benefits experienced by patients who did not have a pneumothorax.The zephyr ebv system ifu and pulmonx training program both specifically reference pneumothorax as a known side effect of this procedure and the published guidelines (valipour, arschang, et al."expert statement: pneumothorax associated with endoscopic valve therapy for emphysema-potential mechanisms, treatment algorithm, and case examples." respiration 87.6 (2014): 513-521).The reported event aligns with the experience observed in the liberate clinical study and is an expected side effect to the zephyr valve treatment.
|
|
The subject had four zephyr valves placed in the left lower lobe (lll) on (b)(6) 2024.On (b)(6) 2025 a secondary bronchoscopy-1 was performed where 4 index valves were removed and 5 valves were implanted.Post-procedure the patient demonstrated acute respiratory acidosis and hypercapnic respiratory failure with tachypnea, tachycardia, and hypercapnea.A chest x-ray showed there was a large right pneumothorax where a chest tube was inserted and the patient's tachycardia, tachypnea, and hypercapnea rapidly resolved.Respiratory acidosis was resolved on (b)(6)2025.On (b)(6) 2025 a secondary bronchoscopy-2 was performed where one valve was removed due to the worsening airleak/pneumothorax.A ct was taken that showed segmental pulmonary embolism where the patient was started on anticoagulation with heparin infusion which was supratherapeutic during monitoring.This led to hemoptysis that improved with cessation of anticoagulation where hemoptysis was resolved on (b)(6) 2025.All valves were removed on (b)(6)2025 with rapid resolution of the leak.The chest tube was removed on (b)(6) 2025 and the patient was discharged on (b)(6) 2025 with resolution of the pulmonary embolism.
|