Model Number 1-7333-75
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/15/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product involved in the report has not been returned.A review of the device history record is in-progress.All information reasonably known as of 21 mar 2025 has been included in the health authority report.Should additional information be obtained a follow-up health authority report will be provided.The information provided by airlife.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to airlife.Airlife has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the airlife complaint database and is identified as complaint - 11747.This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a airlife product is defective or causes serious injury.
 
Event or Problem Description
It was reported that an endotracheal tube cuff leaked which resulted in an unsecured airway, necessitating reintubation.
 
Event or Problem Description
It was reported that an endotracheal tube cuff leaked which resulted in an unsecured airway, necessitating reintubation.
 
Additional Manufacturer Narrative
The product involved in the report has not been returned.A review of the device history record is in-progress.All information reasonably known as of 21 mar 2025 has been included in the health authority report.Should additional information be obtained a follow-up health authority report will be provided.The information provided by airlife.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to airlife.Airlife has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the airlife complaint database and is identified as complaint(b)(4).This information is submitted pursuant to 21 cfr 803, in compliance with the medical device reporting requirement and should not be considered to be an admission that a airlife product is defective or causes serious injury.Complaint was considered not confirmed based on customer narrative of events and provided video.A 24 month review was conducted and 0 additional complaints were identified related to this issue.No trend was observed.The lot number was provided, and the vendor was notified of the incident.This incident was not identified as an increasing trend.No further action to be taken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUNMED
Common Device Name
E/T TUBE CUFFED 7.5MM 30FR
Manufacturer (Section D)
SUNMED HOLDINGS LLC
2710 northridge dr nw suite a
grand rapids MI 49544
Manufacturer (Section G)
SUNMED AP, LTD.
6-9 zhufeng rd., qiannan indus
qianwu town, doumen district
zhuhai guangdong, china 51917 0
CH   519170
Manufacturer Contact
melissa brickley
2710 northridge dr nw suite a
grand rapids, MI 49544
6162598415
MDR Report Key21670719
Report Number1314417-2025-00009
Device Sequence Number8120142
Product Code BTR
UDI-Device Identifier10889483001813
UDI-Public(01)10889483001813(10)2308012011
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K042683
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1-7333-75
Device Catalogue Number1-7333-75
Device Lot Number2308012011(
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2025
Supplement Date Received by Manufacturer02/20/2025
Initial Report FDA Received Date03/21/2025
Supplement Report FDA Received Date05/22/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-