Additional information was received, and section h should be reported as: the manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap/ bipap, and mechanical ventilator devices.The manufacturer received information alleging visualization of particles.No other clinical information or medical interventions were reported.The device was returned to the third-party service center for further evaluation.During the evaluation of the device at the third-party service center, the device was visually inspected and found no evidence of foam degradation, secondary findings were observed.The device's downloaded logs were reviewed by the manufacturer.There was no error code found.The third-party service center concludes that they could not confirm the customer's allegation and there was no visible foam degradation and unit got corrected.In box h: evaluation method code, evaluation results code, component code and conclusion code grid has been updated.
|