• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. VANGUARD SERIES A STANDARD PATELLA 31MM; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. VANGUARD SERIES A STANDARD PATELLA 31MM; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Codes Pain (1994); Loss of Range of Motion (2032)
Date of Event 01/11/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent a left knee arthroplasty revision to address pain, instability, loss of range of motion and loosening of femoral and tibial components approximately two (2) years post-operatively.All components were revised.Attempts have been made, however, no additional information is available.
 
Additional Manufacturer Narrative
(b)(4).D10 - concomitant devices - vanguard posterior stabilized femoral component left 70mm catalog #: 183132 lot #: j6151967, biomet cruciate tibial tray 75mm catalog #: 141234 lot #: j6453739, vanguard e1 posterior stabilized tibial bearing 71/75mm x 10mm catalog #: ep-183640 lot #: 178120.The complainant has indicated that the product will not be returned to zimmer biomet for investigation, as the device was discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Medical records provided confirmed the reported event.The device history records were reviewed and no discrepancies were identified.A definitive root cause could not be determined with the information provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VANGUARD SERIES A STANDARD PATELLA 31MM
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key21675715
Report Number0001825034-2025-00783
Device Sequence Number15531441
Product Code JWH
UDI-Device Identifier00880304436794
UDI-Public(01)00880304436794(17)231217(10)645120
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K040770
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/17/2023
Device Model NumberN/A
Device Catalogue Number184704
Device Lot Number645120
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/27/2025
Supplement Date Received by Manufacturer07/16/2025
Initial Report FDA Received Date03/24/2025
Supplement Report FDA Received Date07/21/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/17/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexMale
Patient Weight89 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-