Model Number LOGIC FEMORAL PS CEM RIGHT SZ 2
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Discomfort (2330); Limb Fracture (4518)
Date of Event 02/25/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.D10: (b)(6) 02-010-01-0320 - logic femoral ps cem right sz 2.(b)(6) 02-012-45-2020 - lgc tibial fit tray cem sz 2f / 2t.
 
Event or Problem Description
As reported, approximately 13 years and 7 months post the initial right total knee arthroplasty, the patient underwent a liner and patella exchange.The patient's main complaint was discomfort and pain around the patella.Exchange of the tibial liner was more because the surgeon was in there.There was no sign of osteolysis.There was no reported breakage of a device or surgical delay/prolongation.The patient was last known to be in stable condition following the event.
 
Additional Manufacturer Narrative
The revision reported may have been due the reported discomfort and pain.The reason for the pain could not be conclusively determined but may be related to the periprosthetic fracture and the observed patella wear that occurred from over 13 years of use.The reason for the fracture could not be determined, but it may have been from an unreported trauma or patient-related condition.However, these cannot be confirmed since the device was not available for evaluation.D4: corrected.H6: corrected investigation clinical codes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOGIC FEMORAL PS CEM RIGHT SZ 2
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key21675880
Report Number1038671-2025-01639
Device Sequence Number20885600
Product Code JWH
UDI-Device Identifier10885862001214
UDI-Public10885862001214
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K033883
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/02/2016
Device Model NumberLOGIC FEMORAL PS CEM RIGHT SZ 2
Device Catalogue Number02-010-01-0320
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/26/2025
Supplement Date Received by Manufacturer04/17/2025
Initial Report FDA Received Date03/24/2025
Supplement Report FDA Received Date04/17/2025
Date Device Manufactured04/04/2011
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Hospitalization;
Patient Age66 YR
Patient SexFemale
-
-