| Catalog Number |
51-100040 |
| Medical Device Problem Code |
Manufacturing, Packaging or Shipping Problem (2975)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that when inspecting the product, it was found with inner packaging damage.There was no patient involvement.There is no additional information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: 51-145140 item name tprlc xr mp t1 pps 14x113mm lot # 7284521 51-107140 item name tprlc 133 mp type1 pps ho 14.0 lot # 7011645 51-149050 item name tprlc xr mp fp t1 pps 5x95mm lot # 7288002 51-103130 item name tprlc 133 t1 pps so 13x146mmtp lot # 7338447 51-105160 item name tprlc xr t1 pps 16x152mm lot # 3795962 g2: foreign: japan the product has been received by zimmer biomet.The investigation is currently in process.Once the investigation has been completed, a follow-up report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4;g3;h2;h3;h6.Product was returned and evaluated.Visual evaluation of the returned product/provided photos identified damage to the sterile packaging (pouch).Sterility had been compromised.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Reported event is not related to a combination of products; therefore, a compatibility review is not applicable.The reported event did not occur in an operating room or as part of a medical procedure; medical records are not available for review.The condition of the device when it left zimmer biomet is considered conforming to specification.The root cause of the reported event can be attributed to transit damage and a packaging design issue.The reported event has been confirmed by evaluation of the returned product and provided photos.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Search Alerts/Recalls
|