| Model Number |
H74939032502110 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3 - date of event: date of event was approximated to 03/01/2025 as the exact event date was not reported.Good faith effort attempts were made to try and retrieve additional details regarding the reported event, but further information was unable to be obtained.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The 99% stenosed target lesion was located in the mildly tortuous and moderately calcified superficial femoral artery.A 5.0x200x150-hybrid sterling balloon catheter was advanced for dilation.During first inflation at 13 atmospheres for 10 seconds, the balloon burst without reaching the burst rated pressure.The device was removed over the wire and the procedure was completed with another of the same device.There were no patient complications as a result of this event.
|
| |
|