| Model Number |
1400500-08 |
| Medical Device Problem Codes |
Material Rupture (1546); Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was performed to treat a heavily calcified and mildly tortuous lesion in the right coronary artery (rca).After a 5 x 8mm nc trek neo balloon dilatation catheter (bdc) was prepared, with no noted issues, it was advance to the lesion through resistance and attempted to be inflated.However, during the second inflation the balloon was noted to rupture at 8atm.Therefore, the bdc was removed and replaced with another nc trek bdc and the procedure was completed.There were no adverse patient effects nor clinical delay reported.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint history identified no other similar complaints from this lot.Based on the information received, the investigation determined that the reported complaints appear to be related to operational context.There was no leak noted to the balloon during preparation or during the first initial inflation, which suggests a product quality issue did not contribute to the reported difficulties.It was reported by the account that the balloon dilatation catheter (bdc) interacted with the heavily calcified and mildly tortuous anatomy resulting in the reported difficulty advancing the device and subsequently the balloon ruptured during the second inflation.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
| |
|
Search Alerts/Recalls
|