Catalog Number A041122110112
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 02/28/2025
Type of Reportable Event Malfunction
Event or Problem Description
In this event it is reported that a protaper fin.21mm/f1 *not ce* broke during use.The broken part was not retrieved.If the root tip is absorbed 2mm below the broken file, the tooth will be extracted.Requested further information.Reportedly, no injury at this time.
 
Additional Manufacturer Narrative
As a result of this malfunction, the potential for surgical intervention exists to preclude permanent damage to a body structure or permanent impairment of a body function as evidenced by previous reported events with similar files.This event, therefore, is reportable per 21cfr part 803.Investigation results: involved product that broke during use was not returned and cannot be analyzed.Moreover, no unused file is available for evaluation.Nothing unusual to report was found during dhr review (batch #1753498).Root causes are not identified.We will track this kind of event and monitor the trend.

potential root causes may be incorrect technique, overuse (number of uses not communicated), excessive wear, patient condition and behaviour during treatment or material issue (no analysis of the broken fragments possible).In addition, there are further circumstances on the conditions in dentistry (used disinfectants, training/knowledge status), or any other environmental conditions, which are unknown to us and may also have an impact on the reported failure mode.Root causes are not identified.All complaints are monitored through the monthly product surveillance committee and a corrective action could be determined by the committee.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
FILE, PULP CANAL, ENDODONTIC
Manufacturer (Section D)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ  CH-1338
Manufacturer (Section G)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ   CH-1338
Manufacturer Contact
dan eagar
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key21677724
Report Number8031010-2025-00041
Device Sequence Number15531575
Product Code EKS
UDI-Device IdentifierD716PTGRF1211
UDI-PublicD716PTGRF1211
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 03/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberA041122110112
Device Lot Number1753498
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date03/12/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/12/2025
Initial Report FDA Received Date03/24/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Usage of Device A
Patient Sequence Number1
Patient SexMale
-
-