ZIMMER BIOMET, INC. 32MM MOD HEAD COCR -6MM NECK; PROSTHESIS, HIPS Back to Search Results
Model Number N/A
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 02/27/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent an initial procedure.The patient did well for the first few month¿s post operations, however, the patient recently suffered recurrent dislocations.No traumatic event proceeded or resulted from any of these dislocations.A ct was ordered, and the femoral stem was deemed to be in an insufficient degree of anteversion.The stem was changed to a modular arcos to correct the anteversion and the high wall liner was exchanged for a dual mobility.The g7 in situ was retained.No signs of infection.No fractures present.
 
Additional Manufacturer Narrative
(b)(4).D10: 010000925 g7 hi-wall e1 liner 32mm c 7240687.51-117090 tprlc 133 mp type1 bm ho 9.0 66438359.G2: foreign: australia.Proposed code: mechanical (g04) - head.Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.D10: 010000701 g7 bonemaster ltd acet shl 48c 7569976.Visual examination of the provided pictures identified the explanted stem and head, covered with bio debris.The stem component is seen with large tissue mass on the porous coating, with some areas of the porous coating with damage.No other observations can be made for the stem and head using the image alone.No product was returned; further visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: two views of the right hip demonstrate a right total hip arthroplasty with screw fixation of the acetabular cup.Mild radiolucency of the cement hardware interface.Femoral component also demonstrates mild radiolucency of the cement hardware interface.Right total hip arthroplasty with possible evidence of early loosening of the acetabular and femoral components.A definitive root cause cannot be determined.Based on the provided image and mmi, the complaint is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional event information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
32MM MOD HEAD COCR -6MM NECK
Common Device Name
PROSTHESIS, HIPS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key21678480
Report Number0001825034-2025-00806
Device Sequence Number20910746
Product Code JDG
UDI-Device Identifier00887868308314
UDI-Public(01)00887868308314(17)330328(10)65875036
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K942479
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number163667
Device Lot Number65875036
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/27/2025
Supplement Date Received by Manufacturer06/10/2025
Initial Report FDA Received Date03/24/2025
Supplement Report FDA Received Date06/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/28/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11 NARRATIVE.; SEE H11 NARRATIVE.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age56 YR
Patient SexFemale
Patient Weight103 KG
-
-