| Model Number |
EM200 |
| Medical Device Problem Codes |
Unintended System Motion (1430); Loss of Power (1475)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/07/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that device with the report of loss of power and unintended system motion.No patient impact reported.On follow-up it was reported that there was no patient or staff impact.
|
| |
|
Additional Manufacturer Narrative
|
|
H3:product analysis for em200 with serial# (b)(6).:evaluation confirmed overheating.The device was tested and the temperature rise was measured to be 29°f.Rear bearings worn.It was also shaft and cable worn.B3: date of event notification:2025-02-07.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|