| Model Number |
3214 |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Inadequate Pain Relief (2388)
|
| Date of Event |
03/03/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event is estimated.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient experienced ineffective therapy due to a combination of high and low impedances.As a result, surgical intervention was undertaken on (b)(6) 2025 to replace the lead.Therapy was restored post-operatively.
|
| |
|