| Model Number |
420194-13 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The instrument involved in this complaint has not been received and/or tested by failure analysis as of the date of this report.If the product is received and evaluated, a supplemental mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a broken cable with a da vinci product.The customer reported event has no report of patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) received the 8mm mega needle driver instrument to perform failure analysis.The instrument was analyzed and found to have one grip cable broken at the proximal end in the housing.As a result, the cable became loose at the distal end.
|
| |
|