• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ESTABLISHMENT LABS S.A. MOTIVA ERGONOMIX ROUND SMOOTHSILK; MOTIVA IMPLANTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ESTABLISHMENT LABS S.A. MOTIVA ERGONOMIX ROUND SMOOTHSILK; MOTIVA IMPLANTS Back to Search Results
Catalog Number ERSD-425
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 03/05/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported a capsular contracture backer grade iii/iv.
 
Additional Manufacturer Narrative
The most common overall indication for reoperation is capsular contracture.(handel, 2006).Breast prostheses are no different from any foreign material implanted into the human body in the sense of triggering a protective immune reaction from the host.This ¿foreign body response¿ (fbr) is universal and ideally removes or otherwise surrounds the ¿irritant material¿ with fibrous tissue to prevent unwanted immune sequelae.A capsule around a breast implant is, therefore, a necessary mechanism of body defense, but if excessive it can lead to pain and deformity of the breast (steiert, et al., 2013).It has been identified several significant risk factors for capsular contracture, including device features (surface, size), surgical factors (periareolar incision, subglandular placement, antibiotic irrigation), the development of hematoma/seroma, and the use of a surgical bra.(calobrace, 2018).Periprosthetic capsules become pathologically active and undergo a ¿constrictive fibrosis¿ due to retraction of the fibrous tissue, which deforms their contents and impairs the aesthetic outcome, manifested by hardening of variable degree and, in advanced cases, by deformity of the breast.(baker et al., 1990).Per our directions for use, each patient must receive the establishment labs s.A.¿motiva implants®: information for the patient¿ during her surgical consultation, it is the surgeons´ responsibility to ensure that the patient completely understands the information regarding the risks, benefits, and recommendations associated with silicone gel-filled breast implant surgery, as well as the complications typical of any type of surgery.This document is available in motiva® website: ifu.Motiva.Health.The instructions for use in the directions for use were reviewed to determine if there are indications that ensure the prevention and good handling of the product.For this event, it is considered that the information is suitable on the section of surgical precautions as follows: ¿capsular contracture occurs when the capsule tightens and squeezes the implant.This can cause the implant to turn rigid (from slightly firm to quite hard) and the firmest ones can cause varying degrees of discomfort, pain, and palpability.In addition to the firmness, capsular contracture can result in a deformed breast, visible surface wrinkling and/or displacement of the implant.Detection of breast cancer by mammography may also be a more difficult.Capsular contracture may be more common following infection, hematoma, and seroma, and the chance of it happening may increase over time.Capsular contracture is a risk factor for implant rupture, and it is the most common reason for reoperation in augmentation and reconstruction patients.Patients should also be advised that additional surgery might be needed in cases where pain and/or firmness are severe (baker grades iii or iv) and that capsular contracture may happen again after additional surgeries.Correction of capsular contracture may require surgical removal or release of the capsule, or removal and possible replacement of the implant itself." also, a complete review of the dhr was carried out, no deviation was found in the process, so it is established that there is no evidence of a relationship between the event and the manufacturing process.General conclusion: the alleged capsular contracture baker grade iii/iv was confirmed according to the clinical evidence provided.The aetiology of capsular contracture is multifactorial and consisting of patient, type of surgery, prosthetic material used, the implant pocket placement, the incision type and the duration of follow-ups (liu et al., 2015) (headon, kasem & mokbel, 2015).Capsular contracture is a risk associated with breast surgery (handel, 2006) and there is no evidence to suggest a link between this particular implant and/or its manufacturing process and a higher risk of capsular contracture occurrence.The cause of capsular contracture is multifactorial (calobrace, 2018), and we cannot conclude that the reported event was caused by the manufacturing process and/or the product itself.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTIVA ERGONOMIX ROUND SMOOTHSILK
Common Device Name
MOTIVA IMPLANTS
Manufacturer (Section D)
ESTABLISHMENT LABS S.A.
building b-25, coyol free zone
alajuela,alajuela
Manufacturer (Section G)
ESTABLISHMENT LABS S.A.
building b-25, coyol free zone
alajuela,alajuela
Manufacturer Contact
susana hidalgo
building b-25, coyol free zone
alajuela,alajuela 
8003554164
MDR Report Key21683491
Report Number3012883202-2025-00523
Device Sequence Number12686034
Product Code FTR
UDI-Device Identifier07445161010612
UDI-Public(01)07445161010612(11)241118(17)291118
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberERSD-425
Device Lot Number24110984
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/05/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date03/24/2025
Supplement Report FDA Received Date03/27/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/18/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age54 YR
Patient SexFemale
-
-